Distances between optical parts: calculator
A collimated laser beam barely changes over a bench, so most distances between flat optics (mirrors, wave plates, cubes) are free: they are set by the size of the mounts, the table holes and the apertures. Only a few distances are fixed by optics: the lens distances in the double-pass cat's eye, the lens spacing of a telescope, and the length difference between the two arms. The calculators below use standard Gaussian-beam optics (ABCD matrices) and the data-sheet values of our parts; every formula is shown next to its result.
0Which distances are fixed and which are free
| Distance | Set by | Our value | How exact |
|---|---|---|---|
| AOM → cat's-eye lens | lens focal length f | 100 mm | not critical for our small RF scans (calculator 3 gives the walk) |
| cat's-eye lens → retro mirror | f (mirror at the focus) | 100 mm | the critical one: an offset δ makes the return beam converge (calculator 3) |
| telescope lens → lens (beam expander) | f₁ + f₂ | 50 + 100 = 150 mm | set by the output beam staying the same size far away (calculator 2) |
| collection lens → photodiode | the spot must fit the chip | about 40–50 mm | loose: the spot only has to fit 3.6 × 3.6 mm with margin |
| control path − probe path (split → cell) | laser linewidth | within about 0.5 m | calculator 4 |
| between flat optics (mirrors, plates, cubes, isolator) | mount size, table holes, apertures | free | the beam must fit every aperture (calculator 2); mounts must not touch |
| the two alignment irises | accuracy wanted | as far apart as possible | calculator 5 |
1Fibre collimator → beam size
| Result | Value | Formula |
|---|
2Beam along a path: size at every part, clipping, lenses
| Part | Position | Beam there | Passes |
|---|
Edit the path (distances, lenses, apertures)
| Type | Value | Label |
|---|
Transmission of a centred Gaussian beam (1/e² radius w): round aperture of radius a: 1 − exp(−2a²/w²); rectangle: erf(√2 a_x/w) · erf(√2 a_y/w). A round aperture of radius 1.5 w passes 98.9 %, so "aperture at least 1.5 × the beam diameter" is the usual rule. "Max beam" is the largest beam diameter the maker allows (the isolator), not a clear aperture. Distances in the examples are typical, not measured: type in your own.
3Double-pass AOM (cat's eye): the distances that matter
| Result | Value | Formula |
|---|
Why d = f and the mirror at f. The ABCD matrix of the unfolded arm (AOM → d → lens → f + δ → mirror → back) has B = 2(f − d) and C = 2δ/f² for small offsets. B = 0 (d = f) means the returning beam comes back through the same point of the AOM whatever the RF frequency; C = 0 (δ = 0) means it comes back collimated. For our ±200 kHz scan B hardly matters (sub-micrometre walk), so the lens-to-mirror distance is the one to get right: put the mirror where the returning beam has the same size far away as the incoming one (or where the 0th and 1st orders are sharpest on a card). Double-pass practice: Donley et al., Rev. Sci. Instrum. 76, 063112 (2005).
4Control and probe path lengths
| Result | Value | Formula |
|---|
Both beams come from one laser, so its phase noise cancels in the two-photon (EIT) resonance only if both arrive at the same time. A delay τ leaves a relative phase variance 2πΔντ (a Lorentzian, white-frequency-noise laser, as in delayed self-heterodyne theory); the noisy part is spread over about a megahertz, far wider than the kilohertz EIT line, so it only lowers the contrast. Real DFB lasers also have slower (1/f) frequency noise; measure the arm lengths on the bench with a tape.
5Alignment accuracy with two irises
| Result | Value | Formula |
|---|
Geometry only: the further apart the irises, the smaller the angle error. See the alignment question in section 09 for the method.
6What these calculators do not do
- Thin lenses and perfect Gaussian beams only: no aberrations, no lens thickness, no diffraction rings from a clipped beam (the transmission is a first-order estimate).
- The AOM's diffraction efficiency for a given beam size is not modelled: ask the maker for the figure at 795 nm and at our beam size.
- A free-space laser (for example the Yanban option) has an elliptical, astigmatic beam: measure its diameters with the camera and type them in as the start diameter.
- The saturated-absorption (SAS) geometry and the mechanical size of each mount are not calculated: check the mount drawings.
7Sources of the formulas and numbers
- Gaussian beams and ABCD matrices: H. Kogelnik and T. Li, "Laser beams and resonators", Appl. Opt. 5, 1550 (1966); waist after a lens: S. A. Self, Appl. Opt. 22, 658 (1983).
- Collimator beam diameter 4λf/(π MFD) and the maximum waist distance: Thorlabs fibre collimator pages (CFC-11X-B sheet); PM780-HP mode field diameter: Thorlabs PM fibre table.
- AOM: Bragg angle λf/(2v), separation λf/v, rise time ≈ 0.64 D/v (transit time of the sound across the beam); G&H 3080-125 outline drawing 97-01598-01 rev D (4.2 mm/µs, 2.5 × 2 mm aperture, 23/34/65 ns for 125/200/400 µm beams).
- Double-pass AOM: E. A. Donley et al., Rev. Sci. Instrum. 76, 063112 (2005) (NIST PDF).
- Phase noise of a delayed copy of a laser: T. Okoshi, K. Kikuchi and A. Nakayama, Electron. Lett. 16, 630 (1980); L. E. Richter et al., IEEE J. Quantum Electron. 22, 2070 (1986).
- Every function of this page is checked against these values by a test script (32 checks: Thorlabs examples, G&H rise times, ray traces and a numerical integration).